| Horizontal System |
| Sample Rate Range |
500MS/s--1GS/s |
| Waveform Interpolation |
(sin x)/x |
| Record Length |
1M |
| SEC/DIV Range |
4ns/div to 40s/div, in a 2, 4, 8 sequence |
Sample Rate and Delay Time Accuracy
|
±50ppm(at over any ≥1ms time interval) |
| Position Range |
2ns/div to 10ns/div; (-4div x s/div) to 20ms;
|
Delta Time Measurement Accuracy (Full Bandwidth)
|
Single-shot, Normal mode:± (1 sample interval +100ppm × reading + 0.6ns); >16 averages:± (1 sample interval + 100ppm × reading + 0.4ns); Sample interval = s/div ÷ 200
|
| Vertical System |
| Vertical Resolution |
8-bit resolution, all channel sampled simultaneously |
| Position Range |
2mV/div to 10V/div |
| Bandwidth |
60MHz |
| Rise Time at BNC( typical) |
5.8ns |
Analog Bandwidth in Normal and Average modes at BNC or with probe, DC Coupled
|
2mV/div to 20mV/div, ±400mV; 50mV/div to 200mV/div, ±2V 500mV/div to 2V/div, ±40V; 5V/div, ±50V
|
| Math |
+, -, *, /, FFT |
| FFT |
Windows:Hanning,Flatop, Rectamgular,Bartlett,Blackman; 1024 sample point
|
| Bandwidth Limit |
20MHz |
| Low Frequency Response (-3db) |
≤10Hz at BNC |
| DC Gain Accuracy |
±3% for Normal or Average acquisition mode, 10V/div to 10mV/div; ±4% for Normal or Average acquisition mode, 5mV/div to 2mV/div
|
DC Measurement Accuracy, Average Acquisition Mode
|
When vertical displacement is zero, and N ≥16:± (3% × reading + 0.1div + 1mV) only 10mV/div or greater is selected; When vertical displacement is not zero, and N≥16: ± [3% × (reading + vertical position) + 1% of vertical position + 0.2div]; Add 2mV for settings from 2mV/div to 200mV/div; add 50mV for settings from 200mV/div to 10V/div
|
Volts Measurement Repeatability, Average Acquisition Mode
|
Delta volts between any two averages of ≥16 waveforms acquired under same setup and ambient conditions |
| Trigger System |
| Trigger Types |
Edge, Video, Pulse, Slope, Over time, Alternative |
| Trigger Source |
CH1, CH2, EXT, EXT/5, AC Line |
| Trigger Modes |
Auto, Normal, Single |
| Coupling Type |
DC, AC, Noise Reject, HF Reject, LF Reject |
Trigger Sensitivity (Edge Trigger Type)
|
DC(CH1,CH2): 1div from DC to 10MHz;1.5div from 10MHz to 100MHz; 2div from 100MHz to Full; DC(EXT): 200mV from DC to 100MHz; 350mV from 100MHz to 200MHz; DC(EXT/5): 1V from DC to 100MHz;1.75V from 100MHz to 200MHz; AC: Attenuates signals below 10Hz; HF Reject: Attenuates signals above 80kHz; LF Reject: Same as the DC-coupled limits for frequencies above 150kHz; attenuates signals below 150kHz
|
| Trigger Level Range |
CH1/CH2: ±8 divisions from center of screen; EXT: ±1.2V; EXT/5:±6V
|
| Trigger Level Accuracy( typical)Accuracy is for signals having rise and fall times ≥20ns |
CH1/CH2: 0.2div × volts/div within ±4 divisions from center of screen; EXT: ± (6% of setting + 40mV); EXT/5: ± (6% of setting + 200mV);
|
| Set Level to 50%(typical) |
Operates with input signals ≥50Hz |
| Video Trigger |
|
| Video Trigger Type |
CH1, CH2: Peak-to-peak amplitude of 2 divisions; EXT: 400mV; EXT/5: 2V
|
| Signal Formats and Field Rates, Video Trigger Type |
Supports NTSC, PAL and SECAM broadcast systems for any field or any line |
| Holdoff Range |
100ns ~ 10s |
| Pulse Width Trigger |
|
| Pulse Width Trigger Mode |
Trigger when (< , >, = , or ≠); Positive pulse or Negative pulse |
| Pulse Width Trigger Point |
Equal: The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level. Not Equal: If the pulse is narrower than the specified width, the trigger point is the trailing edge. Otherwise, the oscilloscope triggers when a pulse continues longer than the time specified as the Pulse Width. Less than: The trigger point is the trailing edge. Greater than (also called overtime trigger): The oscilloscope triggers when a pulse continues longer than the time specified as the Pulse Width
|
| Pulse Width Range |
20ns ~ 10s |
| Slope Trigger |
|
| Slope Trigger Mode |
Trigger when (< , > , = , or ≠ ); Positive slope or Negative slope |
| Slope Trigger Point |
Equal: The oscilloscope triggers when the waveform slope is equal to the set slope. Not Equal: The oscilloscope triggers when the waveform slope is not equal to the set slope. Less than: The oscilloscope triggers when the waveform slope is less than the set slope. Greater than: The oscilloscope triggers when the waveform slope is greater than the set slope.
|
| Time Range |
20ns ~ 10s |
| Overtime Trigger |
|
| Over Time Mode |
Rising edge or Falling edge |
| Time Range |
20ns ~ 10s |
| Alternative Trigger |
|
| Trigger on CH1 |
Internal Trigger: Edge, Pulse Width, Video, Slope |
| Trigger on CH2 |
Internal Trigger: Edge, Pulse Width, Video, Slope |
| Trigger Frequency Counter |
|
| Readout Resolution |
6 digits |
| Accuracy (typical) |
±30ppm (including all frequency reference errors and ±1 count errors) |
| Frequency Range |
AC coupled, from 4Hz minimum to rated bandwidth |
| Signal Source |
Pulse Width or Edge Trigger modes: all available trigger sources The Frequency Counter measures trigger source at all times, including when the oscilloscope acquisition pauses due to changes in the run status, or acquisition of a single shot event has completed. Pulse Width Trigger mode: The oscilloscope counts pulses of significant magnitude inside the 1s measurement window that qualify as triggerable events, such as narrow pulses in a PWM pulse train if set to < mode and the width is set to a relatively small time. Edge Trigger mode: The oscilloscope counts all edges of sufficient magnitude and correct polarity. Video Trigger mode: The Frequency Counter does not work.
|
| Measurement |
| Cursor Measurement |
Voltage difference between cursors: △V Time difference between cursors: △T Reciprocal of △T in Hertz (1/ΔT)
|
| Auto Measuerment |
Frequency, Period, Mean, Pk-Pk, Cycli RMS, Minimum, Maximum, Rise time, Fall Time, +Pulse Width, -Pulse Width, Delay1-2Rise, Delay1-2Fall, +Duty, -Duty, Vbase, Vtop, Vmid, Vamp, Overshoot, Preshoot, Preiod Mean, Preiod RMS, FOVShoot, RPREShoot, BWIDTH, FRF, FFR, LRR, LRF, LFR, LFF |